Kontaktaufnahme

Live-Chat mit Tektronix-Vertretern. Verfügbar von 9 bis 17 Uhr CET Geschäftstage.

Anrufen

Kontaktieren Sie uns telefonisch unter

Verfügbar von 9 bis 17 Uhr CET Geschäftstage.

Download

Laden Sie Handbücher, Datenblätter, Software und vieles mehr herunter:

DOWNLOADTYP
MODELL ODER SCHLÜSSELWORT

Feedback

When acquiring data from one analog input channel with my KUSB module the data looks good, however, when acquiring more than one channel the data looks corrupt, especially when using a high gain. Why is this occurring?

Frage:

When acquiring data from one analog input channel with my KUSB module the data looks good, however, when acquiring more than one channel the data looks corrupt, especially when using a high gain. Why is this occurring?

Antwort:

Product(s):

USB Modules: KUSB-3100 Series: KUSB-3102, KUSB-3108, KUSB-3116,


Question:I get good readings when acquiring from one channel using either the KUSB-3102, KUSB-3108, or KUSB-3116. However, when acquiring more than one channel the data is corrupt, especially when using a high gain. Why is this occurring?

 


Answer:These modules have high impedance inputs (100M ohm).If your signals are coming from a high impedance source, they will not have time to settle out when doing a channel scan.The settling time increases when a high gain is applied.As a remedy, use instrumentation amplifiers between the inputs and our module. Analog Devices part #AD623 is a proven resolution.

 