FAQs – Häufig gestellte Fragen

FAQ FAQ-ID Sort ascending
Model 2520 Remote Test Head

The remote head is available separately.
The number is
2520-300G5
But has to be calibrated with the mainframe..


Diese FAQ bezieht sich auf:
2520.
773561
What is the difference between the P7513A and P7513?

The difference between these two probes is a longer cable on the P7513A.  The added length


Diese FAQ bezieht sich auf:
P7513, P7513A
773521
How do I set up a network drive on my 5 Series MSO or 6 Series MSO?

Introduction


Diese FAQ bezieht sich auf:
MSO54, MSO56, MSO58, MSO58LP, MSO64
773296
Tektronix Video Test Signal generators – and future GPS roll-over dates.

In short, the GPS roll-over date will have no effect until possibly 2029, depending on the product. For more detail Download File.

 

 


Diese FAQ bezieht sich auf:
SPG8000, SPG8000A, TG700, TG8000
773206
Host ID for SourceXpress or for the optional application plug-ins?

Within SourceXpress, from the Tools menu, click on “Licenses” to display the license management screen.  The Host ID is listed at the top of the screen.


Diese FAQ bezieht sich auf:
ENVFL-SS01, ENVNL-SS01, HSSFL-SS01, HSSNL-SS01, MTONEFL-SS01, MTONENL-SS01, OPTICALFL-SS01, OPTICALNL-SS01, PRECOMFL-SS01, PRECOMNL-SS01, RFGENFL-SS01, RFGENNL-SS01
772691
TRCP0300, TRCP0600, TRCP3000 power supply

Tektronix does not offer a replacement part number.


Diese FAQ bezieht sich auf:
TRCP0300, TRCP0600, TRCP3000
772586
Kickstart 2.x.x Software License Installation Process

If you ordered KICKSTARTFL-BASE that is used with the Kickstart 2.0 or higher software application.  If you ordered KICKSTARTFL-HRMA that is used with Kickstart 1.9 software application>


Diese FAQ bezieht sich auf:
KICKSTARTFL-BASE, KICKSTARTFL-HRMA
772141
Error message “CvP Tester Didn’t Run” on my 5 Series MSO.

Instruments that have both a Linux drive and a Windows drive should have the same firmware revision on both drives.


Diese FAQ bezieht sich auf:
MSO54, MSO56, MSO58, MSO64
771991
DMM7510 in 4W RTD outputs a square wave signal that affects Fluke Calibrator during calibration.

The reason there is what appears to be a square wave output when the RTD temperature measurement is turned on is because offset compensation (OCOMP) is enabled by default.

Diese FAQ bezieht sich auf:
DMM7510

729751
How to set ACV DB function in Model DMM6500

In the SCPI command set mode, the following commands will get you there:

:SENSE:FUNC 'VOLT:AC'

:VOLT:AC:UNIT DB

:SENSE:VOLT:AC:DET:BAND 300

:SENSE:VOLT:AC:DB:REF 5


Diese FAQ bezieht sich auf:
DMM6500
729491
Downloads
Download

Laden Sie Handbücher, Datenblätter, Software und vieles mehr herunter:

Go to top